Emerson Process Management MLT 1 TV Converter Box User Manual


 
II) V24/RS232/485 Interface - Single Analyzers and Systems
90003752(1) [AK-Commands] 10/98
AK
2 - 71
AQEF – Read command "Cross interference"
To this read command the CO analyzer or the system will send to the TBCC:
The concentration value in ppm determined and stored in the analyzer. This value will
be stored in the device until a new cross interference will be determined by the "SQEF"
command. The TBCC will control, if limits will be exceeded. The TBCC will also start
actions if necessary.
Read command
AQEF Kn
Read of channel n
Code
Response
AQEF 0 XXX
Interference [ppm]
Error status
Code