II) V24/RS232/485 Interface - Single Analyzers and Systems
90003752(1) [AK-Commands] 10/98
AK
2 - 27
SNAB – Control command "Zerogas calibration"
Starting this command the analyzers in a system or the single analyzer will start a zerogas
calibration. The calibration gas flow will start automatically and the calibration procedure
will run. After this procedure will be over the system, the analyzer in a system or the single
analyzer will change to the stand-by mode. The running calibration procedure can be
canceled with the "STBY" command.
Control command
SNAB K0
Function associated to the whole system unit
Code
SNAB K1 Kn
Function associated to channel n
Function associated to channel 1
Code
Response
SNAB 0
Error status
Code