Emerson Process Management MLT 1 TV Converter Box User Manual


 
II) V24/RS232/485 Interface - Single Analyzers and Systems
90003752(1) [AK-Commands] 10/98
AK
2 - 69
AMBU – Read command "Switching values for autoranging"
To this read command the analyzers in a system or the single analyzer will send to the
TBCC the following data for the called channel (device):
The adjusted switching values in ppm for a changing of ranges with autoranging.
Read command
AMBU K0
AMBU Kn
Read of channel n
Code
Response
AMBU 0 M1 xxx XXX M2 yyy YYY Mn zzz ZZZ
Switch on value of range n
Switch off value of range n
Range No. n
Switch on value of range 2
Switch off value of range 2
Range No. 2
Switch on value of range 1
Switch off value of range 1
Range No. 1
Error status
Code
The values will get the same format for the read of single channels.