AKON – Read command "Signal" (measured concentration value)
To this read command the analyzers in a system or the single analyzer will send to the
TBCC the following data:
The corrected signal (concentration value) valid at that moment. Normally the physical
unit is ppm. The value will be limited to six relevant digits, because it is useless to send
gas concentrations in an accuracy less than pars pro mille.
Example for four relevant digits (default: six digits):
measured
conc. [ppm]
sent
conc. [ppm]
123456 123500
12356 12360
1234.4 1234
123.45 123.5
12.56 12.56
1.23 1.23
Cf. the "SFRZ" command !
Read command
AKON K0
Read of the whole system unit
Code
AKON K1 .... Kn
Read of channel n
Read of channel 1
Code