Agilent Technologies 16760A TV Converter Box User Manual


 
123
Chapter 4: Using the Logic Analyzer in Eye Scan Mode
Setting Up and Running Eye Scan Measurements
memory device versus a memory controller.
Requirements
Qualified eye scans are performed with double edge clocks.
Qualified eye scans are available only in the 800 Mb/s mode of the
16760 logic analyzer.
When the qualifier is in use, there are six channels on the master card
that cannot be measured by eye scan:
Pod 2, clock input (clock K) - the qualifier
Pod 2, channels 14 and 15
Pod 1, channels 0, 1, and 2
This limitation is due to the internal architecture of the analyzer.
You can not perform an eye scan on the J clock of the master logic
analyzer card because this is the target system clock signal. It is used
as the reference signal for the eye scan.
Signal Timing - Burst
Transfers
The qualifier feature is designed for systems using double data rate
bursted transfers. In these systems, data is transferred on each edge of
the clock. In other words, there are two data transfers for each clock
cycle - one on the rising edge and one on the falling edge. Figure 1
shows a burst with four clock cycles and eight data transfers. The clock
cycle begins on the rising edge of the clock.
The direction of the transfer (read versus write) is typically set up via
an earlier command sequence (not shown).