Agilent Technologies 16760A TV Converter Box User Manual


 
233
Chapter 5: Reference
Specifications and Characteristics
400 Mb/s Sampling Mode Specifications and
Characteristics
NOTE: All specifications are marked by "*" (asterisk).
Maximum data rate: 400 Mb/s
Minimum clock interval (active edge
to active edge)*: 2.5 ns
Minimum state clock pulse width: 1.5 ns
Clock periodicity: Periodic or Aperiodic
Number of clocks/qualifiers: 1
Clock polarity: Rising, falling, or both
Minimum data pulse width: 1.5 ns
Number of Channels
With time tags: 34 x (number of modules) - 17
Without time tags: N/A
Maximum channels on single time
base and trigger: 153 (5 modules)
Maximum memory depth: 32 M samples
Setup and Hold Time Range
(manual adjustment),
single edge: +4.5 ns / -2.0 ns to -2.0 ns / +4.5 ns
multiple edge: +5.0 ns / -2.0 ns to -1.5 ns / +4.5 ns
Window width (manual adjustment)*,
single edge: 2.5 ns
multiple edge: 3.0 ns
Window width after running eye finder: 1.25 ns
Adjustment resolution: 100 ps
Time tag resolution: 4 ns
Maximum time count between states: 17 sec
Trigger resources: 8 patterns (=,/=,<,>,<=,>=),
4 ranges (in range, not in range),
2 occurrence counters,
4 flags
Trigger actions: Go to, trigger and fill memory
Maximum trigger sequence levels: 16
Maximum trigger sequencer speed: 400 MHz
Store qualification: Default
Maximum global counter: N/A
Maximum occurrence counter: N/A
Maximum pattern/range term width: 32 bits
Timer value range: N/A
Timer resolution: N/A
Timer accuracy: N/A
Timer reset latency: N/A
Data in to BNC port out: 150 ns
Flag set/reset to evaluation: N/A
Note 2: The resolution of the hardware used to assign
time tags is 4 ns. Times of intermediate states
are calculated.