Chapter 5: Reference
The Eye Scan Tab
Select None De-selects all channels.
See Also “To select channels for the eye scan” on page 119
Scan Settings Subtab
Scan Range Lets you choose from Coarse, Medium, or Fine time and
voltage settings.
The Default button returns to the defaults for the selected
scan range option.
Time Lets you enter the minimum and maximum times (relative
to clock signal edges) and the size of the time windows for
the scan.
Voltage Lets you enter the minimum and maximum voltages
(relative to the threshold voltage) and the size of the time
windows for the scan.
If the Scan voltage over entire range of signal activity
option is selected, the voltage range is determined
automatically during the measurement.
Run Mode Lets you look at eye scan results with demo data or in