Agilent Technologies 16760A TV Converter Box User Manual


 
160
Chapter 5: Reference
The Sampling Tab
Sampling Positions Tab
The Sampling Positions display is a digital "eye" diagram in that it
represents many samples of data captured in relation to the sampling
clock. The transitioning edges measured before and after the sampling
clock result in a picture that is eye-shaped.
You should have already specified the logic analyzer threshold voltage,
but you may adjust it to maximize the width of the measured stable
regions.
Eye finder measures the location of the stable region boundaries and
places the logic analyzer's sampling position in the center of the stable
region.
File menu Lets you save/load eye finder data.
EyeFinder menu Lets you run eye finder, choose the run mode, and access
the “Eye Finder Advanced Settings Dialog” on page 170.
Results menu Let you expand/collapse the signals in a label, set the bus