Chapter 3: Using the Logic Analyzer in Timing or State Mode
Setting Up Triggers and Running Measurements
page 72
“Saving/Recalling Trigger Setups” on page 90
Using Other Trigger Features
Other subtabs in the Trigger tab let you do things like: specifying
whether a state or time count is stored with samples, or setting up the
default storing options.
“To count time, states, or turn counting off” on page 75
“To specify default storing” on page 76
“To specify whether default storing is initially on or off” on page 78
To count time, states, or turn counting off
You can specify whether a time or state count is stored with samples.
1. In the Trigger tab's Settings subtab, select the Count option button and
choose either Off, Time, or States.
The States option is only available in the state sampling mode's 200 Mb/s /
32M State configuration.
2. If you chose States:
a. Select the Define button.
b. In the State count qualify dialog, select the Count if or Count if NOT
option.
c. Specify events that identify the states to be counted or not counted.
d. If you would like to specify the evaluation order of the event list, select
Group events. Then, in the Group Events dialog, either select the Add
parens button to group events or select the Remove parens button to
ungroup events. When you're done grouping events, select the OK
button.
In the state mode's 200 Mb/s / 32M State configuration, memory depth
is reduced by half if all pods are used and time or state counts are being
stored. In all other state mode configurations, time counts require one