Chapter 5: Reference
Error Messages
Maximum of 32 Channels Per Label
The logic analyzer can only assign up to 32 channels for each label. If
you need more than 32 channels, assign them to two labels and use the
labels in conjunction.
Must assign another pod pair to specify actions
for flags
In state sampling mode, when there is only one pod pair assigned to an
analyzer, flags are not available. You must assign another pod pair to
the analyzer in order for flags to be available.
In the timing sampling mode, flags are are always available.
See Also “To assign pods to the analyzer” on page 57
Must assign Pod 1 on the master card to specify
actions for flags
When using a 16760A analyzer in 200Mb/s state mode, Pod 1 on the
master card must be assigned in order to add actions for the flags in a
branch action list. To assign the pod, go to the format tab and select the
"Pod Assignment" button. Drag Pod 1 of the master card from the
unassigned pods list to the analyzer assigned list.
No more Edge/Glitch resources available for
this pod pair
This error occurs when you have used more than 2 edges or glitches
per pod pair in the trigger specification.