Agilent Technologies 16760A TV Converter Box User Manual


 
178
Chapter 5: Reference
The Trigger Tab
See Also “Using Trigger Functions” on page 70
“Editing Advanced Trigger Functions (Timing or 200 Mb/s State Only)” on
page 83
General Timing Trigger Functions
When the timing sampling mode is selected, the following General
Timing trigger functions are found in the Trigger Functions tab.
You can edit most of these trigger functions to specify particular
pattern and edge events or boolean combinations of events.
You can use the following types of events in these trigger functions:
patterns, ranges, flags, counters, timers, and Wait for arm in (see “To
cross-trigger with another instrument” on page 112).
You can use the following actions in these trigger functions: counters,
timers, store control (in the 400 MHz / 32M Sample Transitional or
Store Qualified configuration when Transitions are not being stored),
goto (for arbitrary sequencing), and trigger.
You can expand these trigger functions to see how they are
constructed with the underlying advanced trigger functions.
You can break down these trigger functions to directly edit the
underlying advanced trigger functions. Sometimes you need to break
down a trigger function in order to add other event types such as
counters, timers, flags, and arming from other instruments.
Find pattern
Becomes true when a pattern occurs one time.
Find edge
Becomes true when the specified edge occurs in one sample.
Find edge AND pattern
Becomes true when the specified pattern and the specified edge occurs in
one sample.
Find width violation on a pattern/pulse
Becomes true when the width of a pattern violates minimum and