Agilent Technologies 16760A TV Converter Box User Manual


 
185
Chapter 5: Reference
The Trigger Tab
Advanced - pattern1 OR pattern2
Finds either pattern1 or pattern2 or both in a sample. If you set it to look
for more than 1 occurrence, you can specify whether the occurrences are
consecutive or not. You can also add other events, including labels, to be
searched for.
See Also “Editing Advanced Trigger Functions (Timing or 200 Mb/s State Only)” on
page 83 for more information on using the advanced trigger functions.
Turbo State Trigger Functions
When the state sampling mode's 400 Mb/s / 32M State configuration is
selected, the following Turbo State trigger functions are found in the
Trigger Functions tab.
You can edit these trigger functions to specify particular pattern and
edge events or boolean combinations of events.
You can use the following types of events in these trigger functions:
patterns, ranges, flags, and Wait for arm in (see “To cross-trigger with
another instrument” on page 112).
These trigger functions provide goto and trigger actions. You cannot
modify these actions or insert any other actions.
Find pattern n times
Becomes true when the specified pattern occurs the specified number of
times.
Find pattern1, or reset on pattern2
Becomes true when the specified pattern1 occurs. If pattern2 occurs
before pattern1, then evaluation resets (starts over).
16760 State Trigger Functions
When the state sampling mode's 800 Mb/s / 64M State configuration is
selected, the following 16760 State trigger functions are found in the
Trigger Functions tab.
You can edit these trigger functions to specify particular pattern events
or boolean combinations of events.