Agilent Technologies 16760A TV Converter Box User Manual


 
85
Chapter 3: Using the Logic Analyzer in Timing or State Mode
Setting Up Triggers and Running Measurements
2. Enter an occurrence count value.
3. If the occurrence count is greater than 1, select whether the event should
occur consecutively or eventually.
The event must occur the specified number of times before the action
is taken.
To insert a store action (state mode)
You can insert store actions to override the default storage qualifier for
the samples that cause actions to occur, and you can insert store
actions to turn default storing on or off.
1. In the Trigger tab's Trigger Sequence area, select one of the action buttons
(for example, Trigger or Goto), choose Insert ACTION, choose Store, and
choose either Store sample, Don't store sample, Turn on default
storing, or Turn off default storing.
You can use store actions to set up sequence level storage qualification.
See Also “Storage Qualification” on page 249 in “Understanding Logic Analyzer
Triggering” on page 240
“To specify default storing” on page 76
“To specify whether default storing is initially on or off” on page 78
To insert timer actions/events
Timers are like stopwatches. You can insert actions to start (from
zero), stop (and reset), pause, or resume a timer. You can insert timer
events in a different sequence level to test the value of a timer.
NOTE: No timer is available for the first pod pair assigned to a logic analyzer. For
each additional pod pair assigned to the analyzer, an additional timer is
available.
To insert a timer action
1. In the Trigger tab's Trigger Sequence area, select one of the action buttons
(for example, Trigger or Goto), choose Insert ACTION, choose Timer, and
choose either Start from reset, Stop and reset, Pause, or Resume.