Agilent Technologies 16760A TV Converter Box User Manual


 
181
Chapter 5: Reference
The Trigger Tab
You can expand these trigger functions to see how they are
constructed with the underlying advanced trigger functions.
You can break down these trigger functions to directly edit the
underlying advanced trigger functions. Sometimes you need to break
down a trigger function in order to add other event types such as
counters, timers, flags, and arming from other instruments.
Find pattern n times
Becomes true when the specified pattern occurs in the specified number
of samples (eventually).
Store range until pattern occurs
Becomes true when the specified pattern occurs in the specified number
of samples (eventually) and only stores samples in the specified range
until then.
Store pattern2 until pattern1 occurs
Becomes true when the first specified pattern occurs in the specified
number of samples (eventually) and only stores samples with the second
specified pattern until then.
While storing pattern2, find pattern1
This trigger function has been replaced by the "Store range until pattern
occurs" and "Store pattern2 until pattern1 occurs" trigger functions.
Store nothing until pattern occurs
Becomes true when the specified pattern occurs one time and doesn't
store any samples until then.
Run until user stop
Sets up to never trigger. You must select the stop button to view the
captured data.
Find pattern2 occurring immediately after pattern1
Becomes true when the second specified pattern occurs in the sample
immediately after a sample in which the first specified pattern occurs.
Find pattern1 eventually followed by pattern2