Chapter 5: Reference
Error Messages
No Trigger action found in the trace
specification
This warning occurs when the trigger sequence you specified does not
have at least one trigger and fill memory or trigger and goto action.
The analyzer will still acquire data, but you will need to manually stop
it. For example, when you use the Run until user stop function, to
avoid popping up the status window, set the popup level to errors only.
Slow or Missing Clock
The message "Slow or Missing Clock" only appears in state
measurements. However, if you have another instrument armed by the
state analyzer, a slow or missing clock on the state analyzer will prevent
the other instrument from triggering also.
Possible Causes
Target system is not running properly
Check that the system is running properly. The logic analyzer and other
probing fixtures such as pin extenders can place too much capacitive load
on a system.
Incorrect clock specification
Make sure the device under test clock matches the clock specified under
Sampling.
Also check that the probe's clock channels are attached to the device
under test's clock lines either directly or through an analysis probe.
If you are using an analysis probe, the probe's User's Guide should show
the correct connections and settings.
Bad probe connection
Check that the probe is securely attached to the clock line and is receiving
a signal. The logic analyzer shows activity indicators under the Sampling
and Format tabs.
Incorrect signal level