Konica Minolta CA-100Plus CRT Television User Manual


 
108
Connect the correct probe or expansion
board.
The connected probe or expansion board
differs from the one used on the instrument.
(
The indication in italics shows the model
name of the probe or expansion board.
)
OFFSET ERROR
TOO BRIGHT
NO SYNC. SIGNAL
OVER
SET MAIN PROBE
PROBE ERROR
DATA ERROR
MEMORY ERROR
INCORRECT PROBE
PROBE:
:
INCORRECT BOARD
BOARD:
CA210
CA210
:
Error Message
Cause: (Description) Corrective Action
Zero calibration has not been performed
correctly.
(Zero calibration was performed with insuf-
ficient blocking of entry of light.)
Pressure is given to purobe.
Perform zero calibration again. (Page 34)
(Even if the error message is currently
displayed, measurement will start if the
measuring probe’s receptor is exposed to
light.)
Don’t give the pressure.
Zero calibration is being performed with
insufficient blocking of entry of light.
Block the light completely for all the
measuring probes, and when “DARKEN
PROBE PUSH 0–CAL KEY” appears
press the 0–CAL key again. (Page 34)
Although EXT SYNC mode is selected, the verti-
cal synchronizing signal used for the display is not
input correctly to the terminal on the instrument.
The vertical synchronizing signal used for the dis-
play is outside the range of 40 to 200 Hz.
Input the vertical synchronizing signal correctly.
(When the vertical synchronizing signal is out-
side the range of 40 to 200 Hz/page 28)
Change SYNC mode to NTSC, PAL, UNIV or
INT mode and start measurement.
The measured value is exceeding the
instrument’s measurement range.
The measured value is above 100,000% in
analyzer mode. (Display range over)
Measurement must be performed within
the measuring range.
The measuring probe is not connected to the
probe connector [P1] properly.
Connect the probe to the probe connec-
tor [P1] properly.
(Before connecting/disconnecting the
measuring probe, make sure that the
POWER switch is set to OFF.)
A measuring probe was connected or dis-
connected while the POWER switch was
ON.
Set the POWER switch to OFF first, con-
nect the measuring probe, then set the
POWER switch to ON.
(Before connecting/disconnecting the
measuring probe, make sure that the
POWER switch is set to OFF.)
Measurement is not possible since the mea-
suring circuit is not functioning correctly.
Set the POWER switch to OFF.
If this error still appears even if the
POWER switch is set to ON, the instru-
ment has broken down.
Contact a Konica Minolta authorized ser-
vice facility.
The instrument’s memory is abnormal.
Set the POWER switch to OFF.
If this error still appears even if the
POWER switch is set to ON, the instru-
ment has broken down.
Contact a Konica Minolta authorized ser-
vice facility.
*3
*4
*3
*4
*3
*5
*3
*4