108
• Connect the correct probe or expansion
board.
• The connected probe or expansion board
differs from the one used on the instrument.
(
The indication in italics shows the model
name of the probe or expansion board.
)
OFFSET ERROR
TOO BRIGHT
NO SYNC. SIGNAL
OVER
SET MAIN PROBE
PROBE ERROR
DATA ERROR
MEMORY ERROR
INCORRECT PROBE
PROBE:
:
INCORRECT BOARD
BOARD:
CA210
CA210
:
Error Message
Cause: (Description) Corrective Action
• Zero calibration has not been performed
correctly.
(Zero calibration was performed with insuf-
ficient blocking of entry of light.)
• Pressure is given to purobe.
• Perform zero calibration again. (Page 34)
(Even if the error message is currently
displayed, measurement will start if the
measuring probe’s receptor is exposed to
light.)
• Don’t give the pressure.
• Zero calibration is being performed with
insufficient blocking of entry of light.
• Block the light completely for all the
measuring probes, and when “DARKEN
PROBE PUSH 0–CAL KEY” appears
press the 0–CAL key again. (Page 34)
•
Although EXT SYNC mode is selected, the verti-
cal synchronizing signal used for the display is not
input correctly to the terminal on the instrument.
•
The vertical synchronizing signal used for the dis-
play is outside the range of 40 to 200 Hz.
•
Input the vertical synchronizing signal correctly.
(When the vertical synchronizing signal is out-
side the range of 40 to 200 Hz/page 28)
• Change SYNC mode to NTSC, PAL, UNIV or
INT mode and start measurement.
• The measured value is exceeding the
instrument’s measurement range.
• The measured value is above 100,000% in
analyzer mode. (Display range over)
• Measurement must be performed within
the measuring range.
• The measuring probe is not connected to the
probe connector [P1] properly.
• Connect the probe to the probe connec-
tor [P1] properly.
(Before connecting/disconnecting the
measuring probe, make sure that the
POWER switch is set to OFF.)
• A measuring probe was connected or dis-
connected while the POWER switch was
ON.
• Set the POWER switch to OFF first, con-
nect the measuring probe, then set the
POWER switch to ON.
(Before connecting/disconnecting the
measuring probe, make sure that the
POWER switch is set to OFF.)
• Measurement is not possible since the mea-
suring circuit is not functioning correctly.
• Set the POWER switch to OFF.
If this error still appears even if the
POWER switch is set to ON, the instru-
ment has broken down.
Contact a Konica Minolta authorized ser-
vice facility.
• The instrument’s memory is abnormal.
• Set the POWER switch to OFF.
If this error still appears even if the
POWER switch is set to ON, the instru-
ment has broken down.
Contact a Konica Minolta authorized ser-
vice facility.
*3
*4
*3
*4
*3
*5
*3
*4