Konica Minolta CA-100Plus CRT Television User Manual


 
91
Communications Section
Command
U
A
C
N
Function
Setting user-selected
calibration data or stan-
dard color data (for
memory channel 00)
Sets range of analog
display
Sets memory channel
Inputs ID label
Input format
“U [ x value : y value : Lv value ]”
x value or y value can be input in
any of the three formats given in the
examples below for inputting the
value 0.330:
“0.330”
“.330”
“330”
Spaces must not be input before or
after data.
Lv value
Acceptable range: 0.01 to 999
Number of characters: Up to 4 (in-
cluding decimal point)
Spaces must not be input before or
after data.
a. In xyLv or TuvLv display mode:
“A [ range of x, y :
range of Y ]”
b. In analyzer mode, G standard:
“A [ range of G :
range of R/G, B/G ]”
c. In analyzer mode, R standard:
“A [ range of R :
range of G/R, B/R ]”
Acceptable range: 0.1 to 99
Number of characters: Up to 3 (in-
cluding decimal point)
Spaces must not be input before or
after data.
“C memory channel number ”
Acceptable range: 00 to 10 (If op-
tional Card inserted, acceptable range
becomes 00 to 99.)
“N memory channel number
[ 10-character ID name ]”
or “N [ 10-character ID name ]”
Acceptable range of memory channel
number: 00 to 10 (If optional Card in-
serted, acceptable range becomes 00
to 99.)
Further information
Corresponds to pressing key in
xyLv or TuvLv display mode.
Cannot be used in analyzer mode.
Although would be pressed to
complete input of data if the opera-
tion were performed using keys,
pressing or inputting the cor-
responding command “E” is not
necessary. If “E” is input, the
present measured data will be stored
as the target color data.
This command will not be accepted
unless “I” has been input and zero
calibration has been completed.
Corresponds to pressing RANGE
key.
Although would be pressed to
complete input of data if the opera-
tion were performed using keys,
pressing or inputting the cor-
responding command “E” is not
necessary. If “E” is input, the
present measured data will be stored
as the target color data.
This command will not be accepted
unless “I” has been input and zero
calibration has been completed.
Range data remains in memory even
if CA-100Plus is switched off.
Corresponds to pressing MEMORY
CH or key.
If memory channel number is omit-
ted, memory channel presently be-
ing used will be set.
ID label will remain in memory even
if CA-100Plus is switched off.
Although would be pressed to
complete input of data if the opera-
tion were performed using keys,
pressing or inputting the cor-
responding command “E” is not
necessary. If “E” is input, the
present measured data will be stored
as the target color data.
This command will not be accepted
unless “I” has been input and zero
calibration has been completed.