Konica Minolta CA-100Plus CRT Television User Manual


 
95
Communications Section
2. Data Recalled From Memory
Target color data in memory can be recalled and output using the command “Kxx”, where xx is the memory
channel number from which data is desired. The data which will be recalled and output depends on the present
display mode.
When the optional 4-Probe Expansion Board CA-B04 is used
The probe number for which data will be output should be selected using the command:
“O probe number ”.
In the following, “_” indicates a space.
Format:
CH memory channel number _ P probe number [ serial number (a) ]
[ serial number (b) ] _ standard color data delimiter
1 Memory channel number: Value from 00 to 99; memory channel number from which data is being recalled and
output.
2 Probe number: Value from 1 to 5; probe number for which data is being output.
3 Serial number (a): 8 characters;
For xyLv or TuvLv display mode: Serial number of probe used for calibration to a user-selected reference.
For analyzer (RBG) display mode: Serial number of probe used for setting RGB emission characteristics of
the display.
4 Serial number (b): 8 characters; serial number of probe used for setting target color data.
5 Target color data:
For xyLv or TuvLv display mode:
Format: x value ; y value ; Lv value
x, y values: First three decimal places output.
Lv value: Up to four characters output; if value consists of fewer than four characters, value will be
preceded by the required number of spaces.
For analyzer (RBG) display mode:
Format: R value ; B value ; G value
R, B, G values: Up to five characters output; if value consists of fewer than five characters, value will be
preceded by the required number of spaces.
When the optional 4-Probe Expansion Board CA-B04 is used
If more than one probe number is selected, data will be output consecutively for all selected probe numbers in
numerical order.
Format:
CH memory channel number _ P probe number ..... delimiter
CH memory channel number _ P probe number
.....
delimiter
CH memory channel number _ P probe number
.....
delimiter
Target color data is output according to the format above for the selected display mode.