37
Measurement Preparation
MENU : SELECT
PUSH SPACE KEY
SELECT : M-SPD
AUTO
SELECT : M-SPD
SLOW
SELECT : M-SPD
FAST
CH00 EXT Fd P1
[ ]
Menu selection screen
Measurement speed selection screen
“F” is displayed when the [FAST] was selected.
PQRS
7
TUV
8
WXYZ
MENU
ALPHA
9
GHI
4
JKL
5
MNO LOCK
6
CAL
White
1
ABC
2
DEF
-
SPACE
3
ENTER
Red
0
Green
.
Blue
1
2
3
4
Press the key
until the desired
measurement
speed appears.
[Operating Procedure]
1.Press the key.
The LCD display section will switch to the menu selection
screen.
2. Press the key to open the measurement speed
selection screen.
Each time the key is pressed, the screen will switch in
the order PROBE → SYNC → ID Name input → RANGE
→ Measurement Speed → Number of Digits → RS232C
Baud Rate → PROBE.
3. Press the key to display the desired measure-
ment speed.
Each time the key is pressed, the measurement speed
switches in the order [AUTO] → [SLOW] → [FAST] →
[AUTO].
4. Press the key to confirm the selection.
* By default (factory setting), the instrument is set so that [FAST] will be selected automatically when the POWER switch is set to ON. If
you want to change this setting, refer to page 30.
* To cancel selection of measurement speed, press the
key.
<Notes when Selecting the Measurement Speed>
● The selected measurement speed data will be kept even if the POWER switch is set to OFF.
The selected measurement speed will be effective when the POWER switch is set to ON.