76
CH01 NTSC Ad P1
[ ]
Install the instrument, connect the power cable, and turn ON the power.
Settings section (page 45)
Set up the instrument according to the setting method.
This is not necessary if the instrument has already been set up or if you
are going to perform measurement using Konica Minolta’s calibration
standard and are not going to use the analog display function
Installation/Connection section (page 23)
Measurement Preparation section (page 33)
Perform preparations (instrument setting, zero calibration) that are
required prior to measurement.
Measurement
Before starting measurement, perform the following.
1. Performing Measurement
[Measuring Method]
1.Press the MEMORY CH and keys. keys to select
the memory channel for which user calibration has been
performed (page 50), the RGB emission characteristic for
analyzer mode that has been input (page 58) and the tar-
get color that has been set/changed (page 61).
2. Place the receptor area of the measuring probe flat against
the surface of the display to be measured.
The measurement results will be displayed in the digital and analog dis-
play sections in the selected measurement mode.
<About Low-Luminance Warning>
● The digital and analog display sections are blinking.
• The measured Lv (luminance) is under 0.05 cd/m
2
for white calibration equivalent to Konica Minolta's
calibration standard.
<Notes on Measurement>
● Since the luminance of the display will be unstable for a while immediately after the display is turned ON, the
measured values must be read after they have stabilized.
● Static electricity on the display’s screen surface must be removed as much as possible.
● Perform zero calibration if the ambient temperature has changed.
● When measuring a display at a low luminance level of 0.5 cd/m
2
or less (1.0 cd/m
2
or less when a High Lumi-
nance Measuring Probe(CA-PH02/05) is connected.) for a long period of time, perform zero calibration ap-
proximately every hour.
● Make sure that the measuring probe is placed straight with the display. If it is tilted or moved, accurate measure-
ment cannot be performed.
● Take care not to let the measuring probe be exposed to excessive impact. In addition, do not pull or bend the
cord excessively or exert excessive force on it. Failure to observe these cautions may result in breakdown or
wire-breakage.