92
Command
P
O
L
J
R
G
B
W
Function
Sets probe number for
which data will be dis-
played
Sets probe number for
which data will be out-
put
Sets presently selected
settings as default set-
tings
Sets calibration mode
when CA-100Plus is
set to analyzer (RBG)
display mode
For input of RGB
emission characteris-
tics for display [avail-
able only in analyzer
(RBG) display mode]
Input format
When the optional 4-Probe Ex-
pansion Board CA-B04 is used
“P probe number ”
When the optional 4-Probe Ex-
pansion Board CA-B04 is used
“O probe number ”
Acceptable range: 0 to 5
If data output for more than one probe
connector is desired, more than one
probe number can be input.
For ex-
ample, “0134” will cause data from mea-
suring probes connected to probe connec-
tors 1, 3, and 4 to be output.
For data output from measuring probes
connected to all probe connectors, input
0 as the probe number: “00”
“L”
“J0”: Calibration mode off
“J1”: Calibration mode on
“R”: Sets R emission value
“B”: Sets B emission value
“G”: Sets G emission value
“W”: Sets W (target white value)
Further information
• Probe number to which no probe is
connected cannot be selected.
•
Corresponds to pressing PROBE key.
• Probe number to which no probe is
connected cannot be selected.
• The presently selected probe num-
ber for data output can be determined
by using the command “Z” (status
information request).
• If probe number for which data will
be output is not selected after
POWER switch is first set to ON,
the probe number for data output
will be the same as the probe num-
ber for which data will be displayed.
• Changes the default settings (set-
tings which are automatically set
when POWER is first set to ON) for
the following items:
a. Display mode
b. SYNC mode
c. Memory channel
d. Probe No. for which data will be
displayed. ( When the optional
4-Probe Expansion Board CA-
B04 is used )
• Corresponds to holding pressed
key for five seconds or more.
• If the command “J0” (calibration
mode off) is input after RBGW data
has been set but before the com-
mand “E” (ENTER) is input, the
RBGW data just set will be deleted.
• If the command I, M, A, C, N, or P
is input, calibration mode will au-
tomatically be canceled.
• Corresponds to pressing key
when CA-100Plus is in analyzer
mode.
• This command will not be accepted
unless “I” has been input and zero
calibration has been completed.
• Will be accepted only when calibra-
tion mode has been set (by pressing
key or by inputting the com-
mand “J1”).
• Correspond to pressing
, ,
, and respectively after
key has been pressed in analyzer
mode.