Specifications
Item
Receptor
Measurement area
Detector: Silicon photo cell
φ21 mm
0.01 to 1000 cd/m
2
Displayed in 4 or 3-digit value (Can be chosen)
0.05 to 1000 cd/m
2
±2 %±1 digit of reading(temperature:23˚C±2˚C, relative humidity:(40±10) %)
0.2 %+1 digit (2 σ)
0.05 to 1000 cd/m
2
0.05 to 0.19 cd/m
2
±0.006 for white
0.20 to 0.49 cd/m
2
±0.004 for white
0.50 to 1000 cd/m
2
±0.003 for white
40.00 cd/m
2
±0.002 for white
±0.004 for monochrome *2
0.05 to 0.19 cd/m
2
0.006 (2 σ)
0.20 to 0.49 cd/m
2
0.002 (2 σ)
0.50 to 1000 cd/m
2
0.001 (2 σ)
0.05 to 0.99 cd/m
2
5 measurements/sec. (4.5 measurements/sec.)
1.00 to 1000 cd/m
2
20 measurements/sec. (17 measurements/sec.)
xyLv, XYZ, T∆uvLv, u'v'Lv
RGB
Chromaticity is displayed up to 3 decimal place. (4-digit indication can be chosen.)
∆x∆y∆Lv, R/G B/G ∆G, ∆R B/R G/R (Contrast method)
16 characters by 2 lines (with backlight)
NTSC, PAL, EXT, UNIV, INT
Vertical synchronizing frequency: 40-200 Hz
100 channels
Standard function
RS-232C (38,400 bps or below), USB (1.1 conformity)
Max. 4 points
SDK software (supplied as standard accessory)
Temperature : 10 to 28˚C ; relative humidity 70 % or less with no condensation
Luminance change : ±2 % ±1 digit of reading for white
Chromaticity change : ±0.002 for white, ±0.006 for monochrome from reading of Konica Minolta's
standard CRT, 40.00 cd/m
2
*1
0 to 28˚C ; relative humidity 70 % or less with no condensation
28 to 40˚C ; relative humidity 40 % or less with no condensation
100-240 V , 50-60 Hz, 50 VA
Main body: 340 (W) × 127 (H) × 216 (D), Probe: φ45 × 142
Main body: 3.58 kg, Probe: 285 g
AC power cord (100-120 V or 200-240 V), PC Software for Color Analyzer CA-SDK
4-Probe Expansion Board CA-B04,Measuring Probe(2 m)CA-P02,Measuring Probe(5 m)CA-P05
Display range
Luminance
Chromaticity
Measurement range
Accuracy*1
Repeatability*1
Luminance
Chromaticity
Measurement speed
Display
Measurement range
Accuracy*1
Repeatability
xyLv *3,*4
Digital
Analog
LCD
SYNC mode
Object under measurement
Memory channel
Analyzer function
Interface
Multi-point expansion
Software
Operating temperature/humidity range
Storage temperature/humidity range
Input voltage range
Size
Weight
Standard accessories
Optional accessories
CA-100Plus(Measuring Probe)
*1 : The luminance and chromaticity are measured under Konica Minolta’s conditions (standard CRT(6500 K) is used).
*2 : The luminance for monochrome is measured when the reading of luminance for white is 40.00 cd/m
2
.
*3 : Mesuring probe connected to probe connector P1 only, USB is used(RS-232C;38,400 bps baud rate is used).
*4 : At the CA-200 mode.
• The specifications given in this manual are subject to change without prior notice.
(temperature:23˚C±2˚C,
relative humidity:(40±10) %)