APC AXFA11G TV Converter Box User Manual


 
IM 01E20C01-01E
10-4
10. MAINTENANCE
10.3 Troubleshooting
Although magnetic flowmeters rarely require maintenance, failures may occur when the
instrument is not operated correctly. This section describes troubleshooting procedures
where the cause of the breakdown is identified through receiver indication.
10.3.1 No Indication
Is an error being displayed? Check the converter display.
NO
YES
NO
NO
YES
NO
NO
NO
YES
YES
YES
NO
Is an error being
displayed?
Is anything else
being displayed?
YES
YES
Check whether the flowtube
is filled with fluid.
Is the flowtube
full of fluid?
Remove the tube and
inspect the inside of the pipe.
Is foreign
matter adhering to
the walls?
Does the
arrow on the tube match the
flow direction?
Examine in accordance with
Section 6.5: Alarms.
Is the fuse blown?
Perform either of the following:
a. Reinstall the flowtube in
order to make the arrow and
flow direction match.
b. Reverse the setting for the
flow direction.
Change the piping or the
mounting condition so that
the tube fills with the fluid.
Remove the foreign matter.
START
Contact a service center.
End
Check the resistance between
the flowtube’s A, B, and C
terminals using a multimeter.
Is the
resistance 100 M or more
in each case?
NO
YES
Does the
display return when the fuse
is replaced?
F1005.EPS