6. PARAMETER DESCRIPTION
(3) Display and output condition for setting alarm occurrences
50
Span > 10m/s
51
Span < 0.1m/s
52
TTL>10000p/s
53
TTL<0.0001p/s
54
4-20 Lmt Err
55
Multi Rng Err
56
H/L HH/LL Set
57
Dens Set Err
60
PLS >10000p/s
61
PLS > 5000p/s
62
PLS > 1000p/s
63
PLS > 500p/s
64
PLS > 25p/s
65
PLS > 15p/s
66
PLS > 10p/s
67
PLS > 5p/s
70
PLS<0.0001p/s
71
Meas Mod Set
72
Size Set Err
Closed
(On)
Closed
(On)
Closed
(On)
Closed
(On)
Closed
(On)
Closed
(On)
Closed
(On)
Closed
(On)
Closed
(On)
Closed
(On)
Fixed Stopped Stopped
Alarm Mode
(message)
No
Normal
operation
Stopped
Normal
operation
Alarm Mode
(message)
No
Fixed
Normal
operation
Normal
operation
Alarm Mode
(message)
No
Fixed Stopped Stopped
Alarm Mode
(message)
No
Normal
operation
Normal
operation
Normal
operation
No
Fixed Stopped Stopped
Alarm Mode
(message)
Alarm Mode
(message)
No
Normal
operation
Normal
operation
Normal
operation
Normal
operation
Normal
operation
Stopped
Alarm Mode
(message)
No
Fixed Stopped Stopped
Alarm Mode
(message)
No
Fixed Stopped Stopped
Alarm Mode
(message)
No
Open
(Off)
Fixed Stopped Stopped
Alarm Mode
(message)
No
NO
(G30)
YES
(G30)
Note: • Operations are performed in accordance with above table, when “Open (Off) Act” is set for G20: Alm
Out Act Mode.
• 4-20 mA output upon the occurrence of an alarm will be fixed at the value selected with G21: 4-20mA
Alarm Out.
T0650.EPS
Alarm
record
Alarm description
Selection
(parameter
number)
Alarm
output
4-20 mA
output
Totali-
zation
Pulse
output
Display
unit
Occurring of
any alarm from
50 through 73
Span flow velocity setting
is 11 m/s or more
Span flow velocity setting
is 0.05 m/s or less
Totalization rate is 11000
pps or more
Totalization rate is
0.00005 pps or less
The condition [4-20 low
limit (J11) < 4-20 high
limit (J12)] is not satisfied.
The condition [No. 1 range
< No. 2 range < No. 3 range
< No. 4 range] is not
satisfied for multiple ranges.
The condition [High Alarm
(G11) – Low Alarm (G10) >
H/L Alarm Hys (G14)] or the
condition [High High Alarm
(G13) – Low Low Alarm
(G12) > H/L Alarm Hys
(G14)] is not satisfied.
Mass units have been
selected for Base Flow Unit
(C40) but density is set to 0.
Pulse rate is 11000 pps
or more with 50% duty
selection.
Pulse rate is 10000 pps
or more with 0.05 ms
selection.
Pulse rate is 5000 pps or
more with 0.1 ms selection.
Pulse rate is 1000 pps or
more with 0.5 ms selection.
Pulse rate is 500 pps or
more with 1 ms selection.
Pulse rate is 25 pps or
more with 20 ms selection.
Pulse rate is 15 pps or
more with 33 ms selection.
Pulse rate is 10 pps or
more with 50 ms selection.
Pulse rate is 5 pps or
more with 100 ms selection.
Pulse rate is 0.00005 pps
or less.
Measure Mode (C20) is
set to Enhanced DF
without selecting an
optional code HF1 or
HF2.
A value of 3000.1 mm or
more is set for Nominal
Size (C32).
73
Adh Set Err
Alarm Mode
(message)
No
The condition [Level:1<
Level:2<Level:3<Level:4]
is not satisfied for
adhesion diagnostic level.