IM 01E20C01-01E
8-18
8. OPERATION VIA HART COMMUNICATOR
Setting Password
Example: Set the password to “1 2 3 4”
F0821-1.EPS
ABORT ENTER
3. New Password
AXFA11 : FI-1201
Hot key
1 PV Span 1 m/s
2 Wrt Protect Menu
HELP HOME
ENTER
1
2
AXFA11 : FI-1201
Enter new password to
change state of write
protect
HELP DEL ABORT
ENTER
3
Open “Wrt Protect Menu” in the Hot key Menu.
HELP SAVE
4
YZ /
3
Hot key 2. Wrt Protect Menu
Select “Wrt Protect Menu”.
Select “New Password”.
Enter “1 2 3 4” and press ENTER (F4).
Enter “1 2 3 4” again and press ENTER (F4).
Press OK (F4).
5
6
AXFA11 : FI-1201
Wrt Protect Menu
1 Write protect No
2 Enable Wrt 10min
3 New Password
4 Software Seal Keep
ABORT
SAVE
AXFA11 : FI-1201
Wrt Protect Menu
1 Write protect Yes
2 Enable Wrt 10min
3 New Password
4 Software seal Keep
AXFA11 : FI-1201
Re-enter new password
within 30 seconds:
1234
1234
HELP DEL ABORT ENTER
F4
STU
1
VWX
2
YZ /
3
JKL
4
(ENTER)
F4
STU
1
VWX
2
YZ /
3
JKL
4
(ENTER)
VWX
2
F0821-2.EPS
ABORT
ABORT OK
7
AXFA11 : FI-1201
Change to new
password
HELP DEL
F4
(OK)
AXFA11 : FI-1201
It changed the state
of protection related
password.
HELP OK
Press twice, and return to the “Online” menu.
×2
or
SAVE