6. PARAMETER DESCRIPTION
(8) Item H (Menu H): Display Setting items
Menu H contains setting items that are relevant to display on the display unit.
T0609.EPS
Item
Name
Display unit
(BRAIN)
Data range
Display unit
/BRAIN
Default value
(*): Indicated item
Units
R/W
Position
of decimal
point
Description
H10 Display Select1
(DISP SELECT1)
W Flow Rate(%)
Flow Rate
Flow Rate(mA)
Forward Total
Reverse Total
Dif Total
Flow Rate Selects content of the first line
for Display Mode. Linked with
B40.
H20 Display Cycle
(DISP CYCLE)
W200ms
400ms
1s
2s
4s
8s
400ms Selects the display cycle.
H20 Language
(LANGUAGE)
W English
Japanese
French
German
Italian
Spanish
English Selects the language used by the
display unit. Linked with B10.
H11 Display Select2
(DISP SELECT2)
W Off
Flow Rate(%)
Flow Rate
Flow Rate(mA)
Flow Rate(Bar)
Forward Total
Reverse Total
Dif Total
Tag No
Adhesion Check
Communication
Off Selects content of the second line
for Display Mode. Linked with
B41.
H12 Display Select3
(DISP SELECT3)
W ame as H11
(Display Select2)
Off Selects content of the third line
for Display Mode. Linked with
B42.
H60
(SELF CHECK)
RGood
Error
See “6.5 Alarm Functions”.
H00 Display Set
(DISP SET)
(9) Item J (Menu J): Auxiliary Function Setting items
Menu J contains setting items such as the flow direction, rate limits, and low cut.
T0610-1.EPS
Item
Name
Display unit
(BRAIN)
Data range
Display unit
/BRAIN
Default value
(*): Indicated item
Units
R/W
Position
of decimal
point
Description
J00 Aux
(AUX)
J11 4-20mA Low Lmt
(4-20 LOW LMT)
W-20.0 to 100.0 % 1 -20.0% Sets the low limit for current
output
J12 4-20mA High Lmt
(4-20 HI LMT)
W 0.0 to 120.0 % 1 120.0% Sets the high limit for current
output
J20 Flow Direction
(FLOW DIRECT)
W Forward
Reverse
Forward Selects the flow direction.
J21 Rate Limit
(RATE LIMIT)
W0 to10 % 0 5% Sets the level to reduce output
fluctuation.
J10 4-20mA Low Cut
(4-20 LOW CUT)
W%0 to 10 0 0%
Sets the range in vicinity of 0%
within which the current output
will be 4 mA
J23 Pulsing Flow
(PULSING FLOW)
WNo
Yes
No Selects whether pulsing flow is to
be supported.
J24 T/P Damp Select
(T/P DAMP SEL)
W Damping
No Damping
Damping
Selects whether the flow rate value
obtained through damping calculation
for total/pulse or the instantaneous
flow rate value (no damping) for
total/pulse is to be used.
J22 Dead Time
(DEAD TIME)
Ws0 to 15 0 0s
Sets the dead time to reduce output
fluctuation. When “0” is set, rate
limit function is not available.