APC AXFA11G TV Converter Box User Manual


 
IM 01E20C01-01E
8-14
8. OPERATION VIA HART COMMUNICATOR
9
AXFA11 : FI-1201
Enter meter value
1.000000
1.000000
DEL DEL ABORT ENTER
F4
(ENTER)
10
AXFA11 : FI-1201
Scaled output: 1.000
equal readout
device?
1 Yes
2 No
DEL DEL ABORT ENTER
F4
(ENTER)
Voltmeter reading: 1.010
Enter the voltmeter’s reading of 1.010, and press ENTER
(F4).
(The output of the AXFA11 changes.)
Voltmeter reading: 1.000
Because the reading on the voltmeter is 1.000, select “Yes
and press ENTER (F4).
If the reading is not 1.000, select “No”.
Repeat step until the voltmeter reads 1.000 V.
‘1.01’
9
F0816-3.EPS
AXFA11 : FI-1201
Settting fld dev
output to 20mA
HELP SAVE ABORT
OK
11
12
Press OK (F4). The output signal of 100% is output.
Voltmeter reading: 5.210
Enter the voltmeter’s reading of 5.210, and press ENTER
(F4).
Voltmeter reading: 5.000
Select “Yes” and press ENTER (F4).
Returning fld dev to original output” appears.
Press OK (F4).
Press HOME (F3), and return to the “Online” menu.
AXFA11 : FI-1201
Scaled output: 5.210
equal readout
device?
1 Yes
2 No
HELP DEL ABORT
ENTER
13
AXFA11 : FI-1201
Enter meter value
5.000000
5.000000
HELP DEL ABORT ENTER
AXFA11 : FI-1201
NOTE-Loop may be
returned to automatic
control
DEL DEL ABORT
OK
14
‘5.21’
F4
(ENTER)
F4
(ENTER)
F4
(OK)
F3
(HOME)
F4
(OK)
AXFA11 : FI-1201
Adjustment
1 Auto Zero Exe
2 Magflow Zero
3 D/A Trim
4 Scaled D/A trim
HELP SAVE HOME
ENTER
15
IMPORTANT
D/A trim should be executed only at single range mode. If D/A trim is executed at
Bi direction mode, it is possible that the current output becomes 108%.