Configuration
100 UDA2182 Universal Dual Analyzer Product Manual January 2009
6.15 Auto Cycling Configuration
6.15.1 Overview
Auto cycling provides automated timing, control and functionality for the cleaning and
calibration of input probes. Each input PV has a dedicated auto cycle function block.
The input board type and in the case of pH, the PV type, determines the level of auto
cycling capability, as indicated below:
Input Board Type Auto Cycle Operation
Preamp pH Rinse, Auto Buffer Cal 1 (zero offset), Auto Buffer Cal 2 (slope)
pH/ORP, PV Type not
ORP
Rinse, Auto Buffer Cal 1, Auto Buffer Cal 2
pH/ORP, PV type is ORP Rinse Only
Conductivity Rinse Only
DO ppm Rinse, Auto Air Cal
DO ppb Rinse, Auto Air Cal
Auto cycling is supported with setup menus, status displays and operational displays
(Section 5.7) as well as event (Section 5.11) and calibration history logging (Section 11).
6.15.2 Accessing Auto Cycle Menu
• Press
Setup
to display the Main menu.
• Use the keys to select “Auto Cycling” then press
Enter
to enter the sub-
menu:
Auto Cycle 1 or Auto Cycle 2
• Press to highlight the desired menu selection then press
Enter
to display the
group of parameters.
Refer to “Section 6.4.1 – “General Rules for Editing”.