Outputs Calibration
January 2009 UDA2182 Universal Dual Analyzer Product Manual 179
9.2 Output Calibration
Introduction
The UDA2182 is available with two standard and one optional analog outputs. The
output signals can be adjusted to trim the high and low output current or voltage values
over a range of ± 0.4 % of span to compensate for component tolerance variations.
Accessing the Main Calibration Menu and sub-menus
Press
Calibrate
. The Main Calibration Menu will appear.
CALIBRATION
Input Temp Cal
Output Cal
Cal History
Input PV Cal
CALIBRATION
Input Temp Cal
Output Cal
Cal History
Input PV Cal
Use the
keys to highlight the “Output Cal” selection.
Press
Enter
to display the sub-menu for that selection.
Required equipment
Output calibration involves connecting a meter to the Analyzer’s output terminals.
The meter required for output calibration depends on the type of outputs.
• Current outputs: current meter capable of resolving 0.01 mA over the range 0 to 20
mA dc
• Voltage outputs: a 250 ohm ± 0.05 % shunt and a volt meter (capable of measuring 1
to 5 Vdc within 1 mV)
A screwdriver to fit the terminal block screws and the screw securing the terminal board
retainer is also required.
Procedure
To calibrate outputs, follow the procedure described in Table 9-1 Procedure for
Calibrating Analyzer Outputs. The output terminals are inside the case as shown in
Figures 6-1 through 6-6.