8.7.5 Performing Calibration Trim
Introduction
For most applications entering the cal factor for each cell will achieve satisfactory system
performance. However, it is possible to perform a calibration trim procedure in which the
Analyzer and cell combination are used to measure a reference solution of known
conductivity; the reading of the Analyzer is adjusted to match.
The same procedure can be used to adjust the Analyzer’s reading while the cell is in the
process, if a reference instrument is used to determine the process conductivity. In this
case, the process fluid becomes the “reference solution”.
Calibration trim is recommended for acid concentration measurements above 5%.
Materials
To perform calibration trim using a standard reference solution, follow the instructions in
Table 8-9.
Materials required are:
A reference solution of known conductivity near the point of interest, with the
temperature controlled (or measured and compensated) to within
± 1 °C.
Conductivities of potassium chloride solutions are provided in Table 8-8. Solutions
must be prepared with high-purity de-ionized, CO
2
-free water, and dried potassium
chloride.
For acid concentration applications, a certified reagent grade solution with the
temperature controlled.
A container for the reference solution, large enough to immerse the cell to measuring
depth.
De-ionized water to rinse the cell.