Operating the Analyzer
January 2009 UDA2182 Universal Dual Analyzer Product Manual 35
Table 5-6 Selecting the Pharma Test on Display
Press Action
Enter
to access the Pharma Op Panel. A pop-up dialog box will appear:
Stage 1Pharma Test
80.00PCT Warn
Stage 1Pharma Test
80.00PCT Warn
Various parameters appear for each stage:
Parameter Values Visibility
Pharma Test Selection Stage 1
Stage 2
Stage 3
Pure H2O (PhEur only)
Always on
Pct Warning (% at
which the test warn
occurs)
0 to 100 % Stage 1 only
Test µS/cm
0 to 10 (default=10)
Stage 2 only
Test pH
0 to 14 (default=0)
Stage 3 only
Enter
to access the Stage selection.
to change the selection to
Stage 1, Stage 2, Stage 3 or Pure H2O (PhEur only).
Enter
to make the selection permanent.
to select Pct Warning (Stage 1 only)
Enter
to access the Pct Warning Value and allow editing
to select a value
Enter
to make the selection permanent.
to select Test µS/cm (After Stage 2 is selected)
Enter
to access the Test µS/cm Value
to select a value
Enter
to make the selection permanent.
to select Test pH (After Stage 3 is selected)
Enter
to access the Test pH Value
to select a value
Enter
to make the selection permanent.