Input Calibration
January 2009 UDA2182 Universal Dual Analyzer Product Manual 161
Step Action Screen
5
Press
Enter
6
Follow the prompts at the top
and bottom of the screen.
“Place probe in Sample”
The display will show the
conductivity of the reference
solution as measured by the cell
and Analyzer system.
“Press Enter when stable”
7
Once the reading is stable,
Press
Enter
“Change to Sample value”
Use
to change the value to
match the actual conductivity of
the reference solution at its
current temperature.
“Enter to save, Exit to cancel”
8
Press
Enter
This will save the Calibration
Trim Value. If the calibration trim
adjustment is successful, the
calibration menu will again be
displayed.
Return the cell to the process.
Repeat the operation for the
other cell.
9
If the calibration fails, an error
message will be displayed
across the bottom stripe of the
screen.
Error Messages
Refer to Table 12-2.