Configuration
January 2009 UDA2182 Universal Dual Analyzer Product Manual 105
6.16 Variables Configuration
Overview
The Variables menu allows you to configure the values that variables are set to when the UDA is
first powered on.
This group has two selections:
Analog
This selection lets you configure the initial values of the Analog Variables.
Digital
This selection lets you configure the initial values of the Digital Variables.
Accessing Variables Menu
• Press
Setup
to display the Main menu.
• Use the
keys to select “Variables” then press
Enter
to enter the sub-menu.
• Press to highlight the desired menu selection then press
Enter
to display the
group of parameters.
Refer to “Section 6.4.1 – “General Rules for Editing”.
Table 6-18 Variables Configuration
Sub-menu
selection
Parameter Selection or Range
of Setting
Parameter Definition
Anlg Var 1 Init
Anlg Var 2 Init
Anlg Var 3 Init
Analog
Anlg Var 4 Init
-99999.99 to
99999.99
Initial Values of the Analog Variable applied at
power on.
Dgtl Var 1 Init
Dgtl Var 2 Init
Dgtl Var 3 Init
Digital
Dgtl Var 4 Init
Off
ON
Initial Values of the Digital Variable applied at
power on.