Configuration
January 2009 UDA2182 Universal Dual Analyzer Product Manual 63
6.6 Inputs Configuration
Overview
This group lets you select pH/ORP, Preamp pH, Conductivity, or Dissolved Oxygen
Input type and the associated output parameters.
Accessing Inputs Menu
Press
Setup
to display the Main menu.
Use the keys to select “Inputs” then press
Enter
to enter the sub-menus.
Input 1 and Input 2 – Direct pH/ORP, Preamp pH, Conductivity, or Dissolved Oxygen
are available for selection. Select PV type, read the range, select Temp Type, Solution
Temp Compensation, Bias, Failsafe and Filter Time.
For Dissolved Oxygen, also select the Salinity type and Pressure type.
Press to highlight the desired menu selection then press
Enter
to display the group of
parameters.
Refer to “Section 6.4.1 – ”General Rules for Editing”.
Table 6-5 Input Configuration
Sub-menu
selection
Parameter Selection or
Range of
Setting
Parameter Definition
PV Type pH Glass
pH HPW
pH Durafet
(default)
ORP
The PV type determines the numerical format and the
units of measure on the online PV display. Measured
PV is generally displayed in the highest decimal
precision possible to .001 and has a potentially
displayable range of 0.000 to 99999. The exceptions
are dissolved oxygen, pH, ORP and temperature,
which are displayed with fixed decimal precision. PV
Type determines specific ranges.
PV Range 0.0 to 14.0 pH
-1600 to 1600
ORP
Read Only
Temp Input
(ORP only)
Enable
Disable
Enable to allow “Temp Type” selection – see below.
Input 1 or 2
Direct pH
ORP
Temp Type
8550Ω Therm
(default)
1000Ω RTD
Manual
8550Ω Thermistor
1000Ω Resistance Temperature Detector
Manual