108 UDA2182 Universal Dual Analyzer Product Manual January 2009
6.18 Maintenance Configuration
Accessing Maintenance Menu
Press
Setup
to display the Main menu.
Use the
keys to select “Maintenance” then press
Enter
to enter the sub-
menu.
Press to highlight the desired menu selection then press
Enter
to display the
group of parameters.
Press to highlight the parameter selection, then press
Enter
to allow changes.
Refer to “Section 6.4.1 – “General Rules for Editing”.
Table 6-20 Maintenance Configuration
Sub-menu
selection
Parameter Selection or Range of
Setting
Parameter Definition
SW Version* Software version
number
Read Only
*See note at end of table
Language
Language Set EE
English (default)
Italiano
Deutsch
Francais
Espanõl
Multi-language prompts guide the operator step-
by-step through the configuration process
assuring quick and accurate entry of all
configurable parameters.
Select from: English, French, German, Spanish
and Italian (Language Set EE)
.
Language Set RT
English (default)
Pусский
Türkçe
English, Russian and Turkish (Language Set RT)
Language Set PC
English (default)
Polski
Česká
English, Polish and Czech (Language Set PC)
Read Only -
Language Set
EE
RT
PC
Read only language set of the software
EE - English, French, German, Spanish and Italian
RT - English, Russian and Turkish
PC - English, Polish and Czech
System
Mains Freq
60 Hz (default)
50 Hz
This function determines the frequency of AC line
noise suppression for the input ADC circuitry.