Honeywell UDA2182 TV Converter Box User Manual


 
Outputs Calibration
January 2009 UDA2182 Universal Dual Analyzer Product Manual 183
Step Action Screen
12
Use to select
” 4 mA Offset” and repeat the
process.
OUTPUT 1
20mA Offset
4mA Offset
-147
3
Reset 20mA Offs
Reset 4mA Offs
OUTPUT 1
20mA Offset
4mA Offset
-147
3
Reset 20mA Offs
Reset 4mA Offs
13
Press “Enter” to store the 4mA
Offset value.
Press “Exit” to cancel. The
previous value is retained.
14
If the calibration is suspect,
you can reset the 20mA and
4mA Offset and calibrate
again.
To calibrate additional Outputs, repeat the above steps Including
powering down the unit before changing the connections to the
output terminals.
When output calibration has been completed, re-install the field
wiring removed in step 5.
Disconnect power before opening
the case.
Close the case and power up the unit. Do not apply power until
the case is closed.