Configuration
January 2009 UDA2182 Universal Dual Analyzer Product Manual 67
Sub-menu
selection
Parameter Selection or
Range of
Setting
Parameter Definition
Input 1 or
Input 2
Conductivity
For every cell constant the PV type includes selections for both conductivity µS/cm and
conductivity mS/cm.
Conductivity µS/cm displays µS/cm and provides standard range solution type selections:
None, NaCl, Morpholine, HCL, Acid, and NH3.
Conductivity mS/cm displays mS/cm and provides wide range solution type selections: None,
HCl, NaCl, H2SO4, and NaOH.
Upper range limit defaults according to the table below:
For every cell constant the PV type also includes selections for either TDS ppb/TDS ppm or
TDS ppm/TDS ppt.:
TDS ppb/ppm provides standard or wide solution type selections and
TDS ppm/ppt provides standard or wide solution type selections. Solution selections are the
same as above with the exception of None.
Upper range limit defaults according to the table below:
Cell Const 0.01 Cell Const 0. 1 Cell Const 1 Cell Const 10 Cell Const 25 Cell Const 50
0 - 2 µS/cm
displayable to 200
µS/cm
0 - 0.2 mS/cm
0 - 2000 ppb TDS
0 - 200 ppm TDS
0 - 20 µS/cm
displayable to
2000 µS/cm
0 - 2 mS/cm
0 - 20000 ppb
TDS
0 - 2000 ppm TDS
0 - 200 µS/cm
displayable to
20000 µS/cm
0 - 20 mS/cm
0 - 200 ppm
TDS
0 - 20 ppt TDS
0 - 2000
µS/cm
displayable to
99999 µS/cm
0 - 200 mS/cm
0 - 2000 ppm
TDS
0 - 200 ppt
TDS
0 - 20000
µS/cm
displayable to
99999 µS/cm
0 - 500 mS/cm
0 - 10 % conc
displayable to
20%
0 - 20000
µS/cm
displayable to
99999 µS/cm
0 - 1000
mS/cm
0 - 20 % conc