Honeywell UDA2182 TV Converter Box User Manual


 
Operating the Analyzer
34 UDA2182 Universal Dual Analyzer Product Manual January 2009
The UDA also supports Pharma Europa (PhEur) section 2.2.38, which specifies tests for
determining Highly Purified Water which are identical to USP Stages 1, 2 and 3. PhEur
adds a less demanding test for determining Purified Water at the end of the sequence.
5.8.3 Access to Pharma Display
When Pharma is enabled (see Input Configuration – Section 6.6) press
Display
until you see:
1.30
25.7 ºC
PHARMA 1
Stage 1
74.3%
1
3
4
2
The percent of measured
conductivity relative to test limit
Currently select
Pharma Test
Relay 4 Physical State
White – De-energized
Black - Energized
Relay 3 Physical State
White – De-energized
Black - Energized
Relay 2 Physical State
White – De-energized
Black - Energized
Relay 1 Physical State
White – De-energized
Black - Energized
0.966
The measured temperature-
uncompensated conductivity
of the solution in µS/cm
Purified water test
limit in units of
measured value
µS/cm
Pharma
Measurement
Left Bar graph -Measured
solution conductivity scaled by
the parameter values Pharma
High and Pharma Low
Left Pointer - Test Limit scaled
by parameters Pharma High and
Pharma Low
The measured temperature
of solution in system
temperature Units
Right Pointer -
Warning Limit
relative to test limit
Right Bar graph –Graphical percent of
measured conductivity relative to test limit
1.30
25.7 ºC
PHARMA 1
Stage 1
74.3%
1
3
4
2
The percent of measured
conductivity relative to test limit
Currently select
Pharma Test
Relay 4 Physical State
White – De-energized
Black - Energized
Relay 3 Physical State
White – De-energized
Black - Energized
Relay 2 Physical State
White – De-energized
Black - Energized
Relay 1 Physical State
White – De-energized
Black - Energized
0.966
The measured temperature-
uncompensated conductivity
of the solution in µS/cm
Purified water test
limit in units of
measured value
µS/cm
Pharma
Measurement
Left Bar graph -Measured
solution conductivity scaled by
the parameter values Pharma
High and Pharma Low
Left Pointer - Test Limit scaled
by parameters Pharma High and
Pharma Low
The measured temperature
of solution in system
temperature Units
Right Pointer -
Warning Limit
relative to test limit
Right Bar graph –Graphical percent of
measured conductivity relative to test limit
Figure 5-5 Pharma Display screen example
5.8.4 Displays
The upper left portion of the screen shows the measured temperature uncompensated
conductivity of the solution in uS/cm for Stage 1. For stage 2 and 3 it displays the test
conductivity value entered during the measurement in stage 2
The process temperature is shown in the upper right of the screen
The lower right portion of the screen shows the USP stage 1 purified water limit, the
stage 2 limit of 2.1 uS/cm and the stage 3 purified water test limit at the measured pH.
When the Pharma n display window is active the various stages can be accessed through
the pharma op panel. (Table 5-6)