Honeywell UDA2182 TV Converter Box User Manual


 
Temperature Input Calibration
186 UDA2182 Universal Dual Analyzer Product Manual January 2009
10.2 Temperature Input Calibration
Introduction
Temperature Input Calibration lets you monitor a live temperature reading while
continuing to monitor the sample. The currently displayed temperature value can be
edited through a series of prompts on the screen. The temperature offset value is always
displayed in the temperature units selected in the Maintenance setup menu.
Accessing the Main Calibration Menu and sub-menus
Press
Calibrate
. The Main Calibration Menu will appear.
CALIBRATION
Output Cal
Cal History
Input PV Cal
Input Temp Cal
CALIBRATION
Output Cal
Cal History
Input PV Cal
Input Temp Cal
Use the
keys to highlight the “Input Temp Cal” selection.
Press
Enter
to display the sub-menu for that selection.
Procedure
Table 10-1 Procedure for Calibrating the Temperature Inputs
Step Action Screen
1
Press
Calibrate
CALIBRATION
Output Cal
Cal History
Input PV Cal
Input Temp Cal
CALIBRATION
Output Cal
Cal History
Input PV Cal
Input Temp Cal
Use
to select
” Input Temp Cal”
2
Press
Enter
TEMP INPUT CAL
T1 pH/ORP Cal
T2 Conduc Cal
TEMP INPUT CAL
T1 pH/ORP Cal
T2 Conduc Cal