Agilent Technologies 8590 TV Converter Box User Manual


 
ACPCONTM Continuous Sweep Measurement
ACPCONTM
Continuous Sweep Measurement
Changes the spectrum analyzer’s sweep mode to continuous sweep, and then performs the
previous power measurement (occupied bandwidth, adjacent channel, or channel power) at the
end of every sweep.
Syntax
Equivalent
Softkey:
CONT
MEAS
.
Related Commands: ACP, ACPSNGLM, CHP, CONTS, SNGLS, OBW.
Example
OUTPUT
718*"ACP*"
OUTPUT
718;"ACP~ONTM;"
I4qforrn.s
the adjacent channel power
measurenzent.
Places the spectrum analyzer into continuous sweep
mode
so
that the adjacent channel power measurement will be per-
formed at the end of
ever9
sweep.
Description
When ACPCONTM is executed, the numeric or graph results of adjacent channel (ACP), channel
power (CHP), or occupied bandwidth (OBW) measurements are calculated at the end of every
sweep. While in the continuous sweep measurement, you can change the channel spacing
(ACPSP), channel bandwidth (ACPBW), or another spectrum analyzer setting and the numeric
or graph results for the ACP, CHP, or OBW measurement will be automatically updated.
If ACPCONTM is executed after the adjacent channel power extended command (ACPE), only
the trace results for the top 80
dB
are obtained; the numeric or ACP graph results of the ACPE
measurement are not available in continuous sweep mode.
Programming Commands 5-37