Agilent Technologies 8590 TV Converter Box User Manual


 
GDRVSWP Sweep Time for the Time Window
GDRVSWP
Sweep Time for the Time Window
Specifies the sweep time for the time domain window of the gate utility.
Syntax
/-
sweep time
-\
/
XGDRVSWP
Item
Description/Default
Range
Number
Any real or integer number. Default unit is second.
0 to 65 ms.
Option Required: Option 105. Option 101 is recommended.
Initial Value: 0.
Related Commands: GDRVUTIL, ST.
Example
OUTPUT 718
;
“MOV
GDRVSWP,
1MS
;
‘I
Sets the time
o?ek.zy
to 1
rns.
Description
The positions of
the
gate markers and the gate trigger markers are updated to the
new value
of
the sweep time.
Once you enter a value into GDRVSWP, that value is retained until you change it, or execute
DISPOSE ALL. Pressing
CM]
or turning the spectrum analyzer off does not change the
value of GDRVSWP
You can execute the GDRVSWP command two different ways. You can either execute the
GDRVSWP command directly (for example,
“GDRVSWP
1MS
;
‘I)
or use the MOV command
to move the value for the sweep time into the GDRVSWP command (for example,
“MOV
GDRVSWP ,
1MS
;
‘I).
If you use the MOV command, no text is displayed in the active function area
during command execution.
The GDRVSWP command is a gate utility function (see “GDRVUTIL” for more information
about the gate utility functions). Executing a gate utility command without Option 105
installed in the spectrum analyzer generates the COMMAND ERROR: message and an SRQ 140.
Programming Commands 5-211