Agilent Technologies 8590 TV Converter Box User Manual


 
SAVRCLN Save or Recall Number
Description
SAVRCLN is used to save or recall data from spectrum analyzer memory or from a RAM card.
SAVRCLN is useful if you want to write a program that allows the spectrum analyzer operator
to save data in spectrum analyzer memory or on a RAM card. The SAVRCLN command uses
the SAVRCLF flag information, SAVRCLW information, prefix, and mass storage device when
transferring data.
Because the SAVRCLN command initiates the transfer of data, it should be the last command
specified in the sequence to save or recall data.
The sequence to save or recall data is as follows:
1. Specify either a save or recall operation with SAVRCLF.
2. Indicate
the
type of data to be saved or recalled using SAVRCLW.
3. Specify a RAM card or spectrum analyzer memory as
the
mass storage device with MSI.
4. When saving to or recalling from a RAM card, specify
the
prefix to be used with PREFX.
The prefix is ignored when saving or recalling from spectrum analyzer memory.
5. Enter the number to append to the prefix and initiate the data transfer with SAVRCLN.
When saving trace data, amplitude correction factors, or limit-line tables in spectrum analyzer
memory, specify a number within the trace register number range (0 to TRCMEM
-
1). When
saving state data in spectrum analyzer memory, specify a number within the state number
range (1 to
8).
When saving data on a RAM card, the number plus the number of characters in the prefix must
not exceed eight characters.
Note
With the memory card reader, the spectrum analyzer can read from either a
RAM (random-access memory) card or a ROM (read-only memory card). lb
write to a memory card, the memory card must be a RAM card. The spectrum
analyzer cannot write to a ROM card.
Programming Commands 5481