Agilent Technologies 8590 TV Converter Box User Manual


 
Hewlett-Packard and is available for use with the HP 8590 Series spectrum analyzer. An
example of a personality is the HP 85716A CATV system monitor personality. (U)
FAIL:-
_
_
An error was discovered during the power-up check. The 4-digit by lo-digit code indicates
the type of error. Error codes are described in the spectrum analyzer’s service guide. (H)
File type incompatible
Indicates that the selected file is not a display image file. The file name for a display image
file is always preceded by an “i.” (U)
FREQ UNCAL
If the FREQ UNCAL message appears constantly, it indicates a YTO-tuning error. If this
message appears constantly, perform the CAL FREQ routine. FREQ UNCAL appears briefly
during the CAL FREQ routine; this is normal and does not indicate a problem. (U) and (H)
W>
and
V-0
Function not available in current Mode
Indicates that the function that you have selected can only be used with the spectrum
analyzer mode. You can use the
(-1
key to select the spectrum analyzer mode. (U)
Function not available with analog display
Indicates that the function that you have selected is not compatible with the Analog+
display mode. lb use the function, you must first turn off the Analog+ display mode with
ANALOG* ON OFF. (U)
Gate card not calibrated
This message can indicate that either the CAL AMPTD routine need to be performed before
the time-gating functions can be used, or that something was connected to the GATE
TRIGGER INPUT connector during the CAL AMPTD or CAL FREQ
&
AMPTD routines. If
your spectrum analyzer has an Option 105 installed in it, you should ensure that nothing is
connected to the GATE TRIGGER INPUT connector when the CAL AMPTD or CAL FREQ
&
AMPTD routines are performed. (U) and (H)
INTERNAL LOCKED
The spectrum analyzer’s internal trace and state registers have been locked. lb unlock the
trace or state registers, press
SAV
LOCK ON OFF so that OFF is underlined. For remote
operation, use PSTATE OFF. (U)
INVALID ACTDEF:
-
-
-
The specified ACTDEF name is not valid. See the ACTDEF programming command. (U)
INVALID AMPCOR: FREQ
For the AMPCOR command, the frequency data must be entered in increasing order. See the
description for the AMPCOR programming command for more information. (U)
INVALID BLOCK FORMAT: IF STATEMENT
An invalid block format appeared within the IF statement. See the description for the IF
THEN ELSE
ENDIF
programming command for more information. (U)
INVALID CARD
Indicates one of the following conditions: a card reader is not installed, the memory card is
write-protected (check the position of the switch on the memory card), the memory card is a
read-only memory (ROM) card, or a memory card has not been inserted. This message can
also occur if remote programming commands for the memory card capability are executed
with an HP 859OL or HP 8592L that does not have an Option 003. (U)
INVALID CARD: BAD MEDIA
Indicates the formatting routine
(FOIST
CARD
)
for the memory card could not be
A-4 Spectrum Analyzer Error Messages