Agilent Technologies 8590 TV Converter Box User Manual


 
AMB Trace A Minus Trace B
Example 2
10 OUTPUT
718;"IP;SNGLS;"
20 OUTPUT 718;"MOV TRA, 8000;"
30 OUTPUT
718;"VIEW
TRA;"
40 OUTPUT
718;"MOV
TRB,7000;"
50 OUTPUT
718;"VIEW
TRB;"
60 OUTPUT
718;"AMB
ON;"
70 OUTPUT
718;"BLANK
TRB;VIEW
TRA;"
80
END
Example 3
10 OUTPUT
718;"IP;SNGLS;RL
20DB;"
20 OUTPUT
718;"MOV
TRA,5000;"
30 OUTPUT
718;"VIEW
TRA;"
40 OUTPUT
718;"MOV
TRB,4000;"
50 OUTPUT
718;"VIEW
TRB;"
60 OUTPUT
718;"DL
ODB;"
70 OUTPUT
718."BML-"
9
2
80 OUTPUT
718;"AMB
ON;"
Initializes spectrum analyzer
.Each
element of trace A is set to 8000 measure-
ment units, which is equal to 0
dBm.
Places trace A in the view mode.
Each element of trace B is set to 7000 measure-
ment units, which is equal to -10
dBm.
Places trace B in the view mode.
Subtracts trace B from trace A.
The result is displayed at 1000 measurement
units which is equal to -70
dBm.
Initializes the spectrum analyzer and sets the
reference level.
Sets trace A to 5000 measurement units, which
is equal to -10
dBm.
Places trace A in the mew mode.
Sets trace B to 4000
measuremtst
units, which
is equal to -20
dBm.
Sets display line to 0
dBm,
which is at 6000
measurement units.
Subtracts trace B minus display line. Result is
4000
-
6000 = -2000 measurements units (ofl
screen).
Subtracts trace A minus modified trace B (5000
-(-2000) = 7000 or 10
dBm.
Notice that this has
resulted in a subtraction of amplitude in
dBm,
-10 dBm -(-20
dBm)
= 10
dBm.
90 OUTPUT
718;"BLANK
TRB;VIEW
TRA;"
100 END
5-58
Programming Commands