Agilent Technologies 8590 TV Converter Box User Manual


 
MKSS
Marker to Step Size
Sets the center-frequency step-size to the marker frequency.
Syntax
MKSS
Equivalent
Softkey:
MARKER
-2
CF
STEP .
Related Commands: CF, MKA, MKCF, MKD, MKF, SS.
Example
This example measures a harmonic of the CAL OUT signal.
10 DISP "CONNECT THE CAL OUT TO THE INPUT"
20 OUTPUT
718;"IP;SNGLS;CF
300MHZ;SP
20MHZ;TS;"
30 OUTPUT
718;"MKPK
HI;MKSS;MKD;CF
UP;TS;MKPK
HI;"
40
OUTPUT
718;"MKA?;"
50
ENTER
718;Delta,amp
60
OUTPUT
718."MKF'."
*
,
9
70
ENTER
718;Delta-freq
80 PRINT "DIFFERENCE IN AMPLITUDE IS
",Delta,amp,"dB"
90 PRINT "DIFFERENCE IN FREQUENCY IS
",Delta,freq,"Hz"
100
END
Initializes spectrum
an-
alym
activates single-
sweep mode, changes cen-
ter frequency and span,
updates trace.
Places the marker on the
highest point of the trace,
changes the step size to
the marker frequency, ac-
tivates marker delta, in-
crease center frequency,
update trace, places the
marker at highest point
of the trace.
Gets the amplitude of the
marker:
Puts the spectrum ana-
lyzer response in the com-
puter variable, Delta-Amp.
Gets the frequency of the
marker:
Puts the spectrum ana-
lyzer response in the com-
puter variable,
Deltafreq.
Description
Sets the center-frequency step-size equal to the marker frequency. If in the delta mode, the
step size is set to the delta frequency (absolute value).
5-358 Programming Commands