ST
Sweep Time
Specifies the time in which the spectrum analyzer sweeps the displayed frequency range.
Syntax
Item
Description/Default
Rw&
Number
Any real or integer number. Default unit is seconds.
Within the sweep time
range of the spectrum
analyzer.
Equivalent Softkey:
SWF TfME AUTO MAN .
Sweep Time Range in Zero Span: 15
ms
to
100s.
Sweep Time Range in Zero Span, Option 101 only: 20
,us
to 100s.
Sweep Time Range in Non-zero Span: 20 ms to 100 s.
Step Increment: 2, 3, 5, 7.5, 10, 15 sequence.
Prerequisite Commands: ACTDEF or VARDEF when using a user-defined variable.
Related Commands: AUTO, CONTS, HNLOCK, HNUNLK, RB, SNGLS, SP, SRCPSWP, TS.
Example
OUTPUT 718;"ST
IOOMS;"
Sets the sweep
the
to 100 milliseconds.
Description
When used as a predefined variable, ST returns the sweep time as a real number in seconds.
5-526
Programming Commands