Agilent Technologies 8590 TV Converter Box User Manual


 
ANLGPLUS Analog Plus
ANLGPLUS
Analog Plus
Turns on or off the Analog+ display mode.
Syntax
ON
/
OFF
XANLCPLUS
Equivalent
Softkey:
ZWBLOG*
ON OFF .
Required Option: Option 101 or 301.
Preset State: ANLGPLUS OFF
Related Commands:
DOTDENS.
Example
OUTPUT
718;"ANLGPLUS
ON;"
Turns on the Analog+ display mode.
Description
The Analog+ display mode enables the trace display to emulate an analog display. Emulating
an analog display means that a dot density of up to 40 dots per trace element can be obtained
instead of the usual one point per trace element.
Restrictions: Analog+ display mode is only available with sweeptimes greater than 200 ms.
Certain programming commands are not compatible with the Analog+ display mode. You
should not use the Analog+ display mode with the following programming commands:
AMB, AMBPL, AXB, BML, CLRW (TRB or TRC only), DEMOD,
FFT,
FF’IAUTO, FFTCONTS,
FFTSNGLS, LIMITEST, MEANTH, MINH, MKFC, MKPAUSE, MKSTOP, MKTRACK, MXMH,
ONMKR, PLOT, PP, SPZOOM, SRCTKPK, TH, TRDSP, VAVG, and VIEW.
Some programming commands allow you to specify a destination trace for the trace results. If
you are using the Analog+ display mode and you specify trace A as the destination trace, the
trace results are moved into the 401 point trace and not into the dots on the display. The dots
on the display remain unchanged. The programming commands that allow you to specify trace
A as the destination trace are as follows: ABS, ADD, APB, AVG, COMPRESS,
CONCAT,
DIV,
EXP, INT, LOG, MIN, MIRROR, MOD, MOV, MPY, MXM, PDA, PDF, PEAKS, SMOOTH, SQR, SUB,
TRA, TWNDOW, XCH.
Programming Commands 5-65