Agilent Technologies 8590 TV Converter Box User Manual


 
The preamplifier gain is set by using
EXT?BNAL
PkRAMPG
. This message also sets SRQ 110.
UJ)
CAL: DATA NOT STORED CAL AMP NEEDED
The correction factors are corrupt and cannot be stored. You need to perform the CAL FREQ
&
AMPTD routine before trying to store the correction factors. This message also sets SRQ
110.
(U)
CAL: FM SPAN SENS FAIL
The spectrum analyzer could not set up span sensitivity of the FM coil. (H)
CAL: GAIN FAIL
Indicates the signal amplitude is too low during the CAL AMPTD routine. This message also
sets SRQ 110. (H)
Cal harmonic
>
=
5.7
GHz
NOT found
Indicates that the CAL YTF routine for an HP 85953 cannot find a harmonic of the 300 MHz
calibration signal. If this happens, ensure that the CAL OUT connector is connected to the
spectrum analyzer input, perform the CAL FREQ
&
AMPTD routine, and then perform the
CAL YTF routine again. (U) and (H)
CAL: MAIN COIL SENSE FAIL
The spectrum analyzer could not set up span sensitivity of the main coil. If this
message appears, press
[FREQUENCY),
-37,
(KJ
m,
Mom
1 of 4, More 2 of 4,
DEFAI&T
CAL
DATA , and perform the CAL FREQ routine again. (H)
CAL: NBW 200 Hz notch amp failed
Indicates that the 200 Hz resolution bandwidth is not the correct shape for the calibration
routine. (H)
CAL: NBW 200
Hz
notch failed
Indicates that the 200 Hz resolution bandwidth is not the correct shape for the calibration
routine. (H)
CAL: NBW 200 Hz width failed
Indicates that the 200 Hz resolution bandwidth is not the correct bandwidth for the
calibration routine. (H)
CAL: NBW gain failed
Indicates that one of the resolution bandwidths is not the correct amplitude for the
calibration routine. (H)
CAL: NBW width failed
Indicates that one of the resolution bandwidths is not the correct width for the calibration
routine. (H)
CAL:
PASSCODE
NEEDED
Indicates that the function cannot be accessed without the pass code. For the DEFAULT CA
DATA function, the pass code is setting the center frequency of the spectrum analyzer to
-37 Hz. (M)
CAL: RES BW AMPL FAIL
The relative insertion loss of the resolution bandwidth is incorrect.
SRQ 110. (H)
This message also sets
CAL SIGNAL NOT FOUND
Indicates the calibration signal (CAL OUT) cannot be found. Check
that the CAL OUT and
.L
the spectrum analyzer input connectors are connected with an appropriate cable. If the
calibration signal is connected to the spectrum analyzer input but cannot be found, press
(FREQUENCY),
-37,1Hz),
a,
M
ora
2 of 4 , More 2 of 4 ,
IlEFAaT
GAS
DATA . If the
A-2 Spectrum Analyzer Error Messages