LIMIHI Upper Limit
LIMIHI
Upper Limit
Allows you to specify a fixed trace as the upper limit line.
Syntax
LIMIHI
/
h
user-defined trace
/
XLlMlHl
Related Commands:
IP,
LIMIDEL, LIMIFAIL, LIMILO, LIMISEG, LIMITEST, SENTER.
Example
10
OUTPUT
718;"TRDEF
M,ASK,401;”
20
OUTPUT
718;"MOV
M,ASKC1,1001,1000;"
30
OUTPUT
718;"MOV
M~ASK[101,2001,2000;"
40
OUTPUT
718;"MOV
M~ASK[201,300],3000;"
50
OUTPUT
718;"MOV
M,ASK[301,401] ,400O;”
60
OUTPUT
718;"LIMIHI
M-ASK;"
70
OUTPUT
718;"LIMITEST
ON;"
80
OUTPUT
718;"LIMIFAIL?;"
90
ENTER
718;A
100
DISP A
110
END
Dejines
a trace called “M-ASK.
Moves values into sections of the M-ASK
trace.
Spec@.es
M-ASK
as
the upper limit
line
Turns on limit-line testing.
l&sts
if trace A fails limit-line testing.
Displays result of limit-line testing.
Description
Unlike specifying a limit line with LIMISEG, LIMISEGT, SENTER, or SENTERT, the limit line
specified with LIMIHI is
not
updated if the center frequency, frequency span, sweep time, or
reference level are changed.
Note
Executing IP, LIMIDEL, LIMISEG, LIMISEGT, SENTER, or SENTERT will delete
the limit line specified with LIMILO or LIMIHI. Executing LIMILO or LIMIHI
will delete the limit line specified with LIMISEG, LIMISEGT, SENTER, or
SENTERT.
Use LIMITEST ON to display the limit line trace specified by LIMIHI. Use LIMITEST OFF to
blank the limit line trace specified by LIMIHI.
Programming Commands
5-275