Agilent Technologies 8590 TV Converter Box User Manual


 
MKP Marker Position
MKP
Marker Position
Places the active marker at the given x-coordinate.
Syntax
x
coordinate
predefined variable
b
user-defined variable
/
%
predefined function
/
b
trace element
J
XMKP
Item
Number
User-defined variable
Predefined variable
PredefLned function
Trace element
Description/Default
Range
Any valid integer.
-401 to 401.
A variable defined by VARDEF or ACTDEF commands.
Any valid variable
name.
A command that acts as a variable. Refer to Table
5-l.
Function that returns a value. Refer to Table 5-l.
An element of trace A, trace B, trace C, or a user-defined trace.
Prerequisite Commands: ACTDEF or VARDEF when using a user-defined variable.
Related Commands: MKA, MKCF, MKD, MKMIN.
Example
OUTPUT
718;"IP;"
Initializes spectrum
analyzer:
OUTPUT
718;"MKP
100;"
Mioves the active marker to a element 100 of trace A.
Description
If no marker is active, the marker is turned on with preset type (position) and marker is placed
at the given screen position. The marker is placed on the first displayed trace that is found (in
order): trace A, trace B, or trace C.
If the marker delta mode is active, the value of the marker position is relative to the fixed
marker, and therefore MKP can return a negative position.
Note that MKP and MKCF commands perform different functions. MKCF sets the center
frequency equal to the marker frequency and moves the marker to the center of the screen.
MKP places the marker to the position of the element specified.
Programming Commands 5-347